型 号: | |
所属分类: | 实验室气路工程设计安装 |
报 价: | 10000 |
仪器气(qi)(qi)体(ti)管(guan)道工程(cheng)安装公(gong)司(si)集(ji)中(zhong)供(gong)(gong)气(qi)(qi)系(xi)统(tong)作为气(qi)(qi)体(ti)储(chu)存、调压、输送(song)及供(gong)(gong)应具有(you)以(yi)下几个优点(dian)(dian):一(yi)是集(ji)中(zhong)供(gong)(gong)气(qi)(qi)系(xi)统(tong)将(jiang)气(qi)(qi)源(钢瓶(ping))集(ji)中(zhong)在气(qi)(qi)瓶(ping)储(chu)存间,采用(yong)管(guan)路(lu)系(xi)统(tong)将(jiang)气(qi)(qi)送(song)到(dao)所需的工作地点(dian)(dian),再将(jiang)供(gong)(gong)气(qi)(qi)终端进行(xing)分(fen)流。因(yin)此工作地点(dian)(dian)看不到(dao)气(qi)(qi)瓶(ping),显得规范、整洁。二是结构紧凑、安装方便、操作简(jian)单;三是集(ji)中(zhong)供(gong)(gong)气(qi)(qi)系(xi)统(tong)更(geng)容易(yi)进行(xing)监控。因(yin)此采用(yong)集(ji)中(zhong)供(gong)(gong)气(qi)(qi)系(xi)统(tong)进行(xing)供(gong)(gong)气(qi)(qi)更(geng)安全、更(geng)规范、更(geng)可靠。
仪器气体管道工程安装公司集(ji)中(zhong)(zhong)供(gong)气(qi)(qi)(qi)(qi)(qi)系(xi)统作为气(qi)(qi)(qi)(qi)(qi)体储存、调压、输送(song)及供(gong)应具有(you)以下几个(ge)优(you)点(dian):一是集(ji)中(zhong)(zhong)供(gong)气(qi)(qi)(qi)(qi)(qi)系(xi)统将(jiang)气(qi)(qi)(qi)(qi)(qi)源(yuan)(钢瓶)集(ji)中(zhong)(zhong)在气(qi)(qi)(qi)(qi)(qi)瓶储存间,采用管路(lu)系(xi)统将(jiang)气(qi)(qi)(qi)(qi)(qi)送(song)到所需的(de)工(gong)作地点(dian),再将(jiang)供(gong)气(qi)(qi)(qi)(qi)(qi)终端进行分流。因此工(gong)作地点(dian)看不到气(qi)(qi)(qi)(qi)(qi)瓶,显得规范、整洁。二是结构紧凑、安(an)装方便、操作简单(dan);三是集(ji)中(zhong)(zhong)供(gong)气(qi)(qi)(qi)(qi)(qi)系(xi)统更容易进行监控。因此采用集(ji)中(zhong)(zhong)供(gong)气(qi)(qi)(qi)(qi)(qi)系(xi)统进行供(gong)气(qi)(qi)(qi)(qi)(qi)更安(an)全、更规范、更可靠(kao)。科望达(da)主要提供(gong)氧气(qi)(qi)(qi)(qi)(qi)、氢气(qi)(qi)(qi)(qi)(qi)、氦气(qi)(qi)(qi)(qi)(qi)、氮气(qi)(qi)(qi)(qi)(qi)、乙(yi)炔、压缩空气(qi)(qi)(qi)(qi)(qi)等气(qi)(qi)(qi)(qi)(qi)体的(de)集(ji)中(zhong)(zhong)供(gong)气(qi)(qi)(qi)(qi)(qi)系(xi)统设计(ji)及管路(lu)系(xi)统安(an)装。
一、仪器气体管道工程安装公司技术要求
1、气(qi)瓶间(jian)
①气瓶间应采用300mm以上钢筋混凝土(tu)防(fang)爆(bao)墙,遮阳防(fang)晒,并安装防(fang)爆(bao)门,泄露窗。
②室(shi)内地面(mian)应有(you)防(fang)火花、防(fang)静电措施;室(shi)内电气设备应具备防(fang)爆(bao)功能。
③通风良好,室内(nei)应安装排气扇。
④可燃气(qi)(qi)体(ti)钢瓶(ping)(氢气(qi)(qi)、乙炔)与助燃气(qi)(qi)体(ti)(氧气(qi)(qi))应(ying)隔开放置。
⑤气瓶间(jian)要(yao)设(she)置报警系统
2、工作间
①所有(you)管道要(yao)统(tong)筹(chou)规划(hua)、集中布置、方便(bian)安装、便(bian)于操作、检修方便(bian)。
②所有工艺管道要设置切断阀,并根据需要设置压(ya)力表、温度计、流量(liang)计、过(guo)滤器和减压(ya)阀,阀门和仪表要设置在便于观察(cha)和检修的地方(fang)。
③可燃气体管道要(yao)沿墙明线铺设,方便安装(zhuang)和检修。
④使(shi)用钢瓶气(qi)管道(dao)的设备台应离墙(qiang)布置,且留(liu)出(chu)不小于600mm的检修(xiu)空间(jian)。
⑤氢(qing)气管道应(ying)采用焊接连接,不能采用螺纹连接和使用闸阀。
⑥供(gong)气(qi)系统采用二级减压方式(shi)进行供(gong)气(qi),气(qi)体(ti)由15Mpa减到1.5Mpa后再输(shu)入(ru)到各实(shi)验室。
二、仪器气体管道工程用(yong)材
1、管道、球阀、三通和卡(ka)套(tao)均采用316L不(bu)锈(xiu)钢,高压(ya)软管为不(bu)锈(xiu)钢波纹管,在高压(ya)软管的(de)进气(qi)(qi)端(duan)配置单向阀,防(fang)(fang)止(zhi)(zhi)更换(huan)气(qi)(qi)瓶时(shi),软管内的(de)气(qi)(qi)体外泄,又可防(fang)(fang)止(zhi)(zhi)外界的(de)空气(qi)(qi)混入气(qi)(qi)路中,影响气(qi)(qi)体纯(chun)度。
2、管道(dao)系统:所有气(qi)体(ti)管道(dao)均采(cai)用(yong)B*别316L不锈(xiu)钢管,在管路上有过滤杂质和水(shui)分的(de)装置(zhi)。
3、管道(dao)的连接:汇排流、终端均采用卡套(tao)连接,便于减压器(qi)和(he)阀(fa)门的维(wei)护管理(li)。
4、终端:在每台(tai)仪器之前,配置(zhi)截止(zhi)阀和二级减压器,截止(zhi)阀控制每个(ge)气路(lu)的开启和关(guan)闭,减压器用于(yu)显示和调整每个(ge)终端的压力。
5、气体(ti)管道每(mei)间隔1.5米(mi)采用管码(ma)支架固定。
6、整个(ge)气路(lu)安装(zhuang)完毕后,对整个(ge)系统(tong)做压力(li)测(ce)试,压力(li)24小时后无下(xia)降(jiang)视为合格。
![]() |
在实验室气体管路工程安装施工中,常采用集中供气的方式输送高纯气体,有以下优点: 1、保持气体纯度 气瓶均配有冲洗阀,以排除每次更换气瓶时引入的杂质,确保了管路终端气体的纯度。 2、不间断气体供应 气路控制系统可以手动或自动方式在气瓶之间进行切换,以保证气体的连续供给。 3、低压警示 当气压低于警报*,报警装置可自动启动报警。 4、气体压力稳定 系统采用两级减压(一级由供气控制系统调节,二级由使用点的控制阀调节)方式供气,可得到非常稳定的压力。 5、高效率 通过供气控制系统,可充分使用钢瓶中的气体,减少残气余量,降低用气成本。 6、操作简便 所有气瓶均集中在同一位置,减少了搬运安装等操作,更节约时间及成本费用。 7、减少气瓶的租金 采用中央供气系统,可减少对气瓶数量的要求,从而节省气瓶的租用和购买成本。 8、减少分子筛损耗 控制气体纯度可有效地减少数派对分子筛的使用量(节约成本)。 9、无气瓶在实验室中 采用中央供气系统意味着实验室中没有气瓶装备,有如下好处: ---提高安全性。气瓶可能倒地而导致损坏或伤害。 ---提高安全感。气瓶可能导致气体泄漏、火灾等危险情况。 ---节省空间。从实验室移走气瓶可空出更多的实验空间。 10、其它好处 ---降低控制阀的损坏和丢失率。 ---缩短气阀与仪器间的连接管路的长度,更为方便 |