实(shi)验室集中(zhong)供(gong)气(qi),由实(shi)验室外(wai)专用供(gong)气(qi)区管道(dao)引(yin)入。除洁净空(kong)气(qi)由空(kong)气(qi)压(ya)缩系统直接产生外(wai),其余气(qi)体由高压(ya)缸提供(gong)。
每(mei)种气(qi)(qi)体须***有主供(gong)(gong)和(he)备(bei)用供(gong)(gong)气(qi)(qi)瓶,并安(an)装供(gong)(gong)气(qi)(qi)控制的(de)自动切(qie)换(huan)盘,以保证不间断供(gong)(gong)气(qi)(qi)。此外,主控制阀和(he)泄压阀应安(an)装在实验室外。
实验室气(qi)体采用不锈钢管(guan)道(dao)路运输,一(yi)般在(zai)1.5米(mi)以内,须***用支架(jia)固定在(zai)墙上(shang)。在(zai)实验室,所有(you)(you)管(guan)道(dao)都(dou)安装在(zai)天(tian)花(hua)板下(xia),并(bing)沿(yan)墙壁(bi)铺设。所有(you)(you)管(guan)道(dao)都(dou)标有(you)(you)连接气(qi)体。天(tian)然(ran)气(qi)管(guan)道(dao)应每隔1.5米(mi)清楚标记(ji),标明天(tian)然(ran)气(qi)流向。
所有的减压器都需(xu)要连接到一(yi)个(ge)排气(qi)管,从气(qi)体(ti)储(chu)存(cun)区引出。可燃(ran)和氧化性(xing)气(qi)体(ti)排气(qi)管不能(neng)一(yi)起敷(fu)设。燃(ran)气(qi)管道(dao)不得(de)与电缆、导线同架敷(fu)设。
可(ke)(ke)燃气(qi)(qi)体,如乙炔,需(xu)要与(yu)其(qi)他气(qi)(qi)体分开(kai)引入。氢气(qi)(qi)管(guan)(guan)道与(yu)其(qi)他可(ke)(ke)燃气(qi)(qi)体管(guan)(guan)道平行(xing)敷(fu)设时(shi),其(qi)间(jian)(jian)距不应(ying)(ying)小(xiao)于(yu)0.5m,交叉敷(fu)设的时(shi)间(jian)(jian)间(jian)(jian)隔不应(ying)(ying)小(xiao)于(yu)0.25m,氢气(qi)(qi)管(guan)(guan)道应(ying)(ying)位于(yu)各层以上。
压缩(suo)空气(qi)具有过(guo)滤管道中杂质和水的(de)(de)净化装(zhuang)置(zhi)。净化装(zhuang)置(zhi)需(xu)要(yao)并联(lian)并用单(dan)***的(de)(de)阀门隔离,以便于过(guo)滤装(zhuang)置(zhi)的(de)(de)维护(hu)。
高纯气管线(xian)的连接(jie)采用(yong)无缝焊(han)接(jie)。只有(you)当连接(jie)到阀(fa)门(men)或调(diao)节(jie)器时,才能使用(yong)接(jie)头配件。
每个实(shi)验室都(dou)有单***的控制阀、减压阀和压力计。
通(tong)向工(gong)作(zuo)台的(de)输气(qi)管道(dao)应安装单(dan)***的(de)控制(zhi)阀。工(gong)作(zuo)台上(shang)各种气(qi)体(ti)均匀排(pai)放(fang)的(de)控制(zhi)阀。建议在氦气(qi)管道(dao)前安装气(qi)体(ti)净化(hua)装置(zhi)。
每隔1.5米左右,输气(qi)(qi)管(guan)道(dao)(dao)需要支架。另外,根据燃(ran)气(qi)(qi)管(guan)道(dao)(dao)的(de)弯曲半径,设置适当的(de)支架位(wei)置。应(ying)支撑所有弯管(guan)。燃(ran)气(qi)(qi)管(guan)道(dao)(dao)的(de)所有支架应(ying)镀锌(xin)防(fang)腐。
供气(qi)系统(tong)安装完毕后,按(an)要求进行(xing)相应的(de)强度试(shi)验、密封试(shi)验和稳定(ding)性试(shi)验。